Technical Data
| Gas flow control | |
| Weight | 115 kg (without vaporizers or cylinders) |
| Dimensions (H x W x D) | 137 cm x 80 cm x 80 cm |
| Power | 200 W, typical |
| Operating voltage | 100-240 V~, 45-65 Hz |
| Integrated power backup | at least 30 minutes, typical 90 minutes, depending on ventilation mode |
| Ventilator E-Vent® plus | Electrically driven and electronically controlled |
| Operating Modes | Manual, spontaneous, Volume Mode (IPPV), Pressure Mode (PCV), Optional: Pressure Support (PS), Synchronized volume controlled ventilation (SIMV), optional with PS, Synchronized pressure controlled ventilation, optional with PS |
| Pressure limitation PMAx(in Volume Mode) | (PEEP 10) up to 70 hPa |
| Pressure limitation P”SP (in Pressure Mode) | (PEEP 5) up to 70 hPa |
| Trigger | 0.3-15 l/min |
| Tidal volume (in Volume Mode) | 20-1400 ml 5-1400 ml (option) |
| Tidal volume (in Pressure Mode) | 10-1400 ml |
| Breathing frequency (freq.) | 3-80 1/min |
| Inspiration time [T(INSP)] | 0.2-6.7 s |
| Inspiration/Expiration time ratio (l:E) | max. 5:1 to 1:99 |
| Plateau time [T(IP):T(INSP)] | 0-60 % |
| PEEP in Volume Mode | 0-20 hPa (max. PMAx -10 hPa) |
| PEEP in Pressure Mode | 0-20 hPa (max. P,NSP -5 hPa) |
| Fresh-gas flow | 0 and 0.2-18 Umin |
| T (SLOPE) (in Pressure Mode and Pressure Support) | 0-2 s |
| Total system leakage | < 150 ml at 30 hPa (automatic leak test) |
| O2 flow control | Sensitive ORC function: at least 25 Vol.% or 200 ml/min for nitrous oxide (N20) as carrier gas; 02 concentration: 21-100 Vol.% |
| O2 flush | > 35 Umin |
| 02 safety flow | 0-12 l/min |
| External fresh gas outlet | optional |
| Monitoring | lnspiratory and expiratory concentration of 02, N20, CO2 as well as anaesthetic agents (Halothane, lsoflurane, Sevoflurane, Desflurane); Minute volume (MV) and Tidal volume (VT); Breathing frequency; Peak pressure, plateau pressure, mean pressure, PEEP; Patient compliance CPAT; Option: Functional oxygen saturation (SpO2)
The following parameters maybe displayed as waveforms: Concentration of CO2, O2, as well as anaesthetic agents, airway pressure, inspiratory and expiratory flow; Option: Plethysmogram; Bar graph display of volumeter and tidal volume; Virtual flow tubes or O2, AIR, N2O; Display of graphical trends and numerical lists of measured values; AutoSet for alarm limits |
| Serial interface | 3 x RS 232 |
| Protocol | Medibus |
| Absorber volume | 1.5 L |